Print Back

Designed to provide an atmosphere free from air-borne particles, Completely self-contained unit; not dependent on other equipment, functions effectively in uncontrolled areas. Filtered air is directed forward through a HEPA filter on the top of the chamber. 3/4" novaply, white high-pressure plastic laminate on exposed surfaces. Interior surfaces sealed to prevent shedding of particulates. Also available in 16 ga. cold rolled steel, painted white, or stainless steel with 1 /4" clear acrylic side panels. Square CRS, painted support frame. Liberty Dust Guard prefilters, 30% efficient NBS. HEPA filters 99.99% efficient on particles 0.3 micron in size or larger, Gasketing and fastening devices assure a non-leaking filter installation. Unit is standard without a work surface, uses the existing work area. Exceeds Federal Standard 209 and ISO 14644-1. Class 100 (ISO 5) conditions, High capacity forward curved blowers with direct drive blower motor(s), dynamically balanced to maintain constant air flow without the use of speed controls belt adjustment to twice the clean filter resistance. 120V, 1 phase, 60 HZ standard. 90 FPM +/- 20%. Noise level conforms to OSHA noise standards. Available in stainless steel construction, quoted on request.

OPTIONS:

  • Support frame, constructed of 1W x W x 'Xs" wall sq. CRS tubing.
  • Stainless steel supports optional.
  • Choice of exterior colors.
  • Lights-125 ft. candles location.
  • D.O.P Test Port: 1/4" N.P.T. plug to facilitate sampling of upstream concentration.
  • Custom work area heights 1/4" clear acrylic side panels.
  • Work Surface.
Vertical Flow 5010T Series
Unit of Measure
Items Vertical Flow 5010T Series
5-310 T24-36
Vertical Flow 5010T Series
Vertical Flow 5010T Series
5-410 T24-48
Vertical Flow 5010T Series
Vertical Flow 5010T Series
5-510 T24-60
Vertical Flow 5010T Series
Vertical Flow 5010T Series
5-610 T24-72
Vertical Flow 5010T Series
Vertical Flow 5010T Series
5-810 T24-96
Vertical Flow 5010T Series
No. of HEPA's N/A 1 N/A 1 N/A 1 N/A 1 N/A 2
Est. Running Amps N/A 2.7 N/A 2.7 N/A 2.7 N/A 5.4 N/A 5.4
No. of Blowers N/A 1 N/A 1 N/A 1 N/A 2 N/A 2
Dim. A N/A 39.5 Inch N/A 51.5 Inch N/A 63.5 Inch N/A 75.5 Inch N/A 101 Inch
Wt. N/A 550 lbs N/A 600 lbs N/A 700 lbs N/A 800 lbs N/A 1000 lbs
No. of Light/Watts N/A 1/30 N/A 1/40 N/A 1/40 N/A 1/60 N/A 1/80
No. of Blowers Pre-filter N/A 1 N/A 1 N/A 1 N/A 2 N/A 2